User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Sub Fluke 83,85,87 Freq & Duty Cycle (1 year) CAL VER DATE: 26-Sep-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 11 NUMBER OF LINES: 73 CONFIGURATION: HP 3325B STANDARD: Stanford Research FS700 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N K P J F W 1.002 ASK+ X 1.003 STD Stanford Research FS700 1.004 HEAD {* FREQUENCY TEST *} 1.005 JMP 2.001 1.006 EVAL 2.001 DISP Select the 400mV Range on the UUT. 2.001 DISP Press the Hz button to select the Frequency Counter. 2.002 DISP Connect the HP 3325A SIGNAL output to the 2.002 DISP UUT V Ohm Diode input using a 50 Ohm feedthrough 2.002 DISP termination. 2.003 HEAD FREQUENCY TESTS: {19.999kHz Range} 2.004 JMP 3.001 2.005 EVAL 3.001 M3325 0Voff 3.002 3325 19.000kH 0.005% 0.001U 150mV SI L #! Test Tol 1.95, Sys Tol 0.095, TUR 20.5263 (>= 4.00). 4.001 HEAD FREQUENCY TESTS: {199.99kHz Range} 4.002 JMP 5.001 4.003 EVAL 5.001 3325 190.00kH 0.005% 0.001U 300mV SI L #! Test Tol 10.5, Sys Tol 0.95, TUR 11.0526 (>= 4.00). 6.001 HEAD {* FREQUENCY SENSITIVITY AND TRIGGER LEVEL TESTS *} 6.002 JMP 7.001 6.003 EVAL 7.001 DISP Press the Hz button to select Duty Cycle mode. 7.002 JMP 8.001 7.003 EVAL 8.001 3325 100H 425mV SQ S L 8.002 ACC 50.0pct 0.2% 8.003 MEMI Use keyboard to enter actual UUT reading in Percent: 8.004 MEME 8.005 MEMC pct 4U #! Test Tol 4, Sys Tol 0.1, TUR 40.0000 (>= 4.00). 9.001 DISP Press the RANGE button to select the 4V range. 9.002 3325 100H 848mV SQ S L 9.003 ACC 50.0pct 0.2% 9.004 MEMI Use keyboard to enter actual UUT reading in Percent: 9.005 MEME 9.006 MEMC pct 16U #! Test Tol 16, Sys Tol 0.1, TUR 160.000 (>= 4.00). 10.001 DISP Rotate the UUT function switch to VDC. 10.001 DISP Press the Range button to select the 4V range. 10.001 DISP Press the Hz button twice to select Duty Cycle mode. 10.002 M3325 1.6Voff 10.003 3325 100H 3.4V SQ S L 10.004 ACC 50.0pct 0.2% 10.005 MEMI Use keyboard to enter actual UUT reading in Percent: 10.006 MEME 10.007 MEMC pct 14U #! Test Tol 14, Sys Tol 0.1, TUR 140.000 (>= 4.00). 11.001 DISP Rotate the UUT function switch to OFF. 11.002 DISP Remove all connections. 11.003 HEAD 11.004 END #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 0 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.